Bending beam load cell GS-1A

Download PDF datasheet

Bending beam load cells for weighing platforms, hoppers, tanks, etc. of low and medium capacity

Bending beam load cell

OIML C3 homologated from 75 kg to 300 kg

Upon request with ATEX certification

Click on the image to enlarge

Bending beam load cell GS-1A Bending beam load cell GS-1A Drawing
ATEX OIML Analog Output

Technical specifications

MaterialStainless steel
Nominal load10, 15, 20, 25, 30, 50, 75, 100, 150, 250, 300, 500, 750 kg
Limit load150 % F.S.
Breaking load300 % F.S.
Supply voltageVcc 15 Max
Output2 mV/V
Output tollerance0,1 % F.S.
Zero tollerance1 % F.S.
Linearity0,03 % F.S.
Hysteresis0,03 % F.S.
Reapeatability0,03 % F.S.
Input resistance380-386 ohm
Output resistance350-353 ohm
Insulation resistance>= 2000 MOhm
Creep (30 Minutes)0,03 % F.S.
Thermal compensation-10 / +40°C (14 / +104°F)
Operating temperature-20 / +70°C (-4° / +158°F)
Temperature deviation zero0,003 % F.S. / °C
Temperature deviation full scale0,002 % F.S. / °C
Protection classIP 67
Standard cable length5 m – 6 x 0,30 (shield not connected to the load cell body)

For complete and up-to-date technical specifications please refer to the datasheet you can download here

Download PDF datasheet

Connections

TypeSHIELDED CABLE 6C R6 PVC 6X0,30
Length5 m (shield not connected to the cell body)
Red+ IN
Black- IN
Yellow+ REF
Blue- REF
Green+ OUT
White- OUT
Connections Bending beam load cell GS-1A

Load cell mounting

General guidelines for installing and mounting load cells:

  • Respect the direction in which forces are applied to the load cells
  • Do not exceed the temperature limits shown on the rating plate
  • Do not remove the labels from the cells
  • Protect cables properly with sheaths or cable ducts
  • Connect each cell to earth with a suitable cable, bridging the plates of the mounting accessories if necessary
  • Do not weld on the mechanical structure once the cells are in place
  • Avoid or minimise mechanical constraints between the fixed structure and the weighed structure, to prevent weighing errors and zero return problems
  • When vibrations or dynamic loads are present, protect the cells with anti-vibration mounts
  • Protect the cells from dust build-up with suitable covers